Beamline Layout

The beamline has four different parts:
(1) The 2m-long undulator with a period of 32.9mm and 60 periods
(2) The front-end with beam-transport, filters, white-beam slit systems, and shutters
(3) The beamline optics like monochromators, mirrors, and other components
(4) The end-station with sample environment and detectors

(1) The period of the undulator was determined by the requested large overlap of the first and third harmonic in order to avoid changes in the harmonic or an intensity variation of more than a factor of 5 over the range of an EXAFS-scan.

Existing undulator in the Max-von-Laue Hall. The undulator for P64 has a similar design.

Undulator chamber installed in the beam and labeled with many warning signs since it is quite delicate.

(2) The filter unit of P65 has four different filters: glassy carbon with thicknesses of 0.4, twice 1mm in different units, and 4mm.
The slit-system consists of three different pairs of slits: 2 vertical and 1 horizontal pairs. The first vertical pair has a minimum opening of 1mm, the other two pairs can be completely closed.

The first girder with beam diagnosis tools.

The white-beam slit system and the fast valve.

The filter-unit of P65, the one for P64 is hidden behind it.

The safety shutters for P64 (background) and P65 (foreground) and additional shielding.

(3) The optics hutch is the first enclosure in the experimental hall. In the case of P64, two different monochromators are installed for separate purposes: The first one is a channel-cut monochromator for fast energy scans within 50ms, and the second one is optimized for conventional energy scans in the minute time-scale with a fixed exit. Two mirrors reduce higher harmonic radiation, and they can be used to focus the beam.

P64 Optics hutch

The new high-heatload fixed-exit monochromator.

(4) The experimental hutch is the place, where users install the sample and collect data. The sample can be a simple metal foil, and the data are collected in transmission mode with two ion-chambers, or it can be a complex system in a specialized sample-environment like a furnace with different gas-inlets, a mass-spectrometer for measurements of reaction-output, and a fluorescence detector in order to measure low-concentration samples.

3D Sketch of the experimental and control hutches

The new 100-pixel Ge detector at the old A1 hutch during the site-acceptance test.